4 kft upol - control parameters, Control parameters – Metrohm tiamo 2.3 Manual User Manual

Page 827

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3

■■■■■■■■

811

Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.

Range

0 ... 999999 s

Default value

0 s

5.6.3.5.3.4

KFT Upol - Control parameters

Tab: Method

KFT Upol Properties... Control parameters

Command name

Name of the command.

Input

25 characters

Endpoint

EP with

Measured value for the endpoint.

Range

-200.0 ... 200.0 µA

Default value

25.0 µA

Selection

off

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical, fast titrations.

Advertising