5 met cond - stop conditions, Stop conditions – Metrohm tiamo 2.3 Manual User Manual

Page 698

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

682

■■■■■■■■

tiamo 2.3

The conductivity

measured at a particular temperature is thus auto-

matically converted to the conductivity

of a freely selectable reference

temperature (usually 20 C or 25 C). The conversion takes place with
the aid of the temperature coefficients

with respect to the reference

temperature .

Conductivity at reference temperature

Temperature coefficient

Conductivity at reference temperature

Conductivity at measuring temperature

Measuring temperature

Reference temperature

Temperature coefficients with respect to the reference temperature
(unit: %/°C).

5.6.3.3.6.5

MET Cond - Stop conditions

Tab: Method

MET Cond Properties… Stop condition

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.

Stop volume

Stop when the entered volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Range

0.00000 ... 9999.99 mL

Default value

100.000 mL

Selection

off