8 set ph - additional evaluations, Additional evaluations – Metrohm tiamo 2.3 Manual User Manual

Page 727

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3

■■■■■■■■

711

5.6.3.4.3.8

SET pH - Additional evaluations

Tab: Method

SET pH Properties... Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for the evaluation of titration curves can
be activated and defined on this tab:

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Time or Volume) for the fixed endpoint from the
measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.4.3.10, page 713).

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.4.3.10, page 713).

[Delete]

Delete the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.