6 empty, 1 empty - overview, Empty – Metrohm tiamo 2.3 Manual User Manual

Page 1216

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1200

■■■■■■■■

tiamo 2.3

5.6.6.6

EMPTY

5.6.6.6.1

EMPTY - Overview

Dialog window: Method

EMPTY Properties... EMPTY - 'Command name'

Command for Emptying cylinder and tubing of a dosing unit.

Devices
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 841, 842, 851, 852, 857, 859, 888, 890,
901, 902, 904, 905, 906, 907

Titrino: 736, 751, 758, 799

Dosing Interface: 846

Conductometer: 856

pH/Ion - Meter: 867

Sample Processor: 730, 774, 778, 789, 814, 815, 864, 874

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the command EMPTY are configured in the following
dialog window:

EMPTY - Properties

Command variables
The following command variables are generated in the method run by the
command EMPTY and can be used in formulas under the designation
'Command name.Variable designation':

Identifi-
cation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
not ever been started

.CYL

Cylinder volume of the exchange or dosing unit used for
the command

Advertising