5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

932

■■■■■■■■

tiamo 2.3

Commands
The following standard addition commands can be selected:

STDADD man
Standard addition with manual addition of the standard addition solu-
tion.

STDADD dos
Standard addition with addition of the standard addition solution from
a dosing device.

STDADD auto
Standard addition with automatic addition of the standard addition
solution from a dosing device in such a way that a constant potential
difference results.

5.6.4.3

MEAS pH

5.6.4.3.1

MEAS pH - Overview

Dialog window: Method

MEAS pH Properties.. MEAS pH - 'Command name'

Command for potentiometric pH measurements with pH electrodes.

Instruments
This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 841, 842, 857, 888, 901, 902, 904, 905,
906, 907

Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798,
799

pH/Ion - Meter: 867

Robotic Titrosampler: 855

Appearance
The command has the following appearance:

Parameters
The parameters for the command MEAS pH are configured on the fol-
lowing 4 tabs:

General/Hardware
Parameters for devices, sensors and stirrers.

Measuring parameters
Parameters for setting the measurement procedure.