9 met cond - volume window, 10 met cond - fixed endpoint evaluation – Metrohm tiamo 2.3 Manual User Manual

Page 705

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3

■■■■■■■■

689

5.6.3.3.6.9

MET Cond - Volume window

Dialog window: Method

MET Cond Properties… Conductometric evalua-

tion

Evaluation with volume window (mL) [New]/[Properties] Volume win-

dow #

Volume windows are defined sub-regions (windows) on the volume axis
for which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.

Lower limit

Lower limit for the volume window.

Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Upper limit

Upper limit for the volume window.

Range

0.00000 ... 9999.99 mL

Default value

9999.99 mL

Smoothing

Measure for the curve smoothing used for the data processing. The higher
the smoothing that is selected, the more the measurement curves will be
smoothed.

Range

0 ... 150

Default value

0

5.6.3.3.6.10

MET Cond - Fixed endpoint evaluation

Dialog window: Method

MET Cond Properties… Additional evalua-

tions

Fixed endpoint evaluation [New]/[Properties] Fixed endpoint evalua-

tion #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Advertising