5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1114

■■■■■■■■

tiamo 2.3

Desig-
nation

Description

.ETR

Result of the electrode test:

0 = No data available

1 = Short circuit

2 = Excessive start drift

3 = Wrong buffer

4 = Problems with the diaphragm

5 = Reference system defective

6 = No buffer values for the temperature or excessive devia-
tion between the buffers

7 = Partial short circuit

8 = Glass defective

9 = Poor electrode

10 = Usable electrode

11 = Good electrode

12 = Excellent electrode

.FIN

Command status; 1 = Command has ended at least
once
, 0, invalid (variable not available) = Command has
never ended

.LCO

Loop counter = current quantity of completed runs, both of
Repeat loops and of While loops

.LST

Start time of the loop command (Date/Time)

.SLO

Electrode slope of the sensor used for the command (in mV
for ISE sensors)

5.6.5.8.2

ELT LOOP - Properties

Dialog window: Method

ELT LOOP Properties...

Command name

Name of the command.

Input

25 characters

In this dialog window, the general parameters are set to the buffer used
and to the electrode test.