5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

678

■■■■■■■■

tiamo 2.3

Start volume

Start volume

Volume to be dosed with the indicated dosing rate before the start of the
titration.

Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Dosing rate

Rate (volume/time unit) at which the start volume is to be added. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Pause

Pause

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.

Range

0 ... 999999 s

Default value

0 s

5.6.3.3.6.4

MET Cond - Titration parameters

Tab: Method

MET Cond Properties… Titration parameters

Command name

Name of the command.

Input

25 characters

Parameters defining the run of the titration.

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for the Measured value
acceptance
and Dosing of increments of these sets are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.

Selection

slow | optimal | fast | User

Default value

optimal