5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

690

■■■■■■■■

tiamo 2.3

Measured value
Range

-2000.0 ... 2000.0 mS/cm

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.4

SET

5.6.3.4.1

SET - Overview

Menu item: Method

Insert New command...

Command for Set endpoint titrations.

Principle
With this command titrations to one or two preset endpoints are carried
out. The termination of the titration at the endpoint is either drift-control-
led or after a waiting period. The volume that has been added when the
endpoint is reached gives the calculable reagent consumption.

Reagent addition takes place in three phases during the titration:

Initial dosing
During this phase the dosing rate increases continuously. It starts with
the Min. rate and increases to the Max. rate.

Continuous dosing
The dosing continues to be carried out at the Max. rate until the
Control range is reached.

Control range
Dosing is finely controlled in this range. Shortly before the endpoint is
reached, dosing is carried out only at the Min. rate (Control range).