The command presets the multimeter as specified range and resolution for AC
voltage measurement, but does not start the measurement.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{200 mV|2 V|20 V

|200 V|750 V|AUTO}

AUTO

<resolution> Discrete

Refer to Table 4-2

1ppm x range


Explanations:
The unit is V.
For <range>: MIN = 0. 2 (200 mV), MAX = 750; DEF=AUTO.
<resolution> relates to both the measurement rate and integration time, see
Table 4-2.



3. CONFigure:CURRent:DC

Command Format:
CONFigure:CURRent:DC
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}

Function:
The command presets the multimeter as specified range and resolution for DC
current measurement, but does not start the measurement.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{200 µA|2 mA|20 mA

|200 mA|2 A|10 A |AUTO}

AUTO

<resolution> Discrete

Refer to Table 4-2

1ppm x range


Explanations:
The unit is A.
For <range>: MIN = 0. 0002 (200Μa); MAX = 10, DEF=AUTO.
<resolution> relates to both the measurement rate and integration time, see
Table 4-2.