[SENSe:]RESistance:RANGe? [MINimum|MAXimum] – Queries the current range
of 2-wire resistance measurement.
Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
setting of resistance.

Return Format:
The query returns the range of 2-wire resistance measurement.


7. [SENSe:]FRESistance:RANGe

Command Format:
[SENSe:]FRESistance:RANGe {<range>|MINimum|MAXimum}
[SENSe:]FRESistance:RANGe? [MINimum|MAXimum]

Function:
The commands set and query the range of 4-wire resistance measurement.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{200 Ω|2 kΩ|20 kΩ|200 kΩ

|2 MΩ|10 MΩ|100 MΩ}

200 kΩ


Explanations:
[SENSe:]FRESistance:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of 4-wire resistance measurement from {200 Ω|2 kΩ|20 kΩ|200 kΩ|2
MΩ|10 MΩ|100 MΩ}; thereinto, MIN = 200 Ω, MAX = 100 MΩ and the unit is Ω. If
a 2 kΩ range is required, enter 2000 as its parameter.
[SENSe:]FRESistance:RANGe? [MINimum|MAXimum] – Queries the current
range of 4-wire resistance measurement.
Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
setting of resistance.

Return Format:
The query returns the range of 4-wire resistance measurement.