Sense:]frequency:voltage:range, Sense:]period:voltage:range – RIGOL DM3058 User Manual

Page 129

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

4-36

8. [SENSe:]FREQuency:VOLTage:RANGe

Command Format:
[SENSe:]FREQuency:VOLTage:RANGe {<range>|MINimum|MAXimum}
[SENSe:]FREQuency:VOLTage:RANGe? [MINimum|MAXimum]

Function:
The commands set and query the voltage range under frequency measurement.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{200 mV|2 V|20 V

|200 V|750 V}

20 V

Explanations:
[SENSe:]FREQuency:VOLTage:RANGe {<range>|MINimum|MAXimum} - Sets
the desired voltage range under frequency measurement from {200 mV|2 V|20
V|200 V|750 V}; thereinto, MIN = 200 mA, MAX = 750 V and the unit is V. If a
200 mV range is required, enter 0.2 as its parameter.
[SENSe:]FREQuency:VOLTage:RANGe? [MINimum|MAXimum] – Queries the
current voltage range under frequency measurement.

Return Format:
The query returns the voltage range under frequency measurement.



9. [SENSe:]PERiod:VOLTage:RANGe

Command Format:
[SENSe:]PERiod:VOLTage:RANGe {<range>|MINimum|MAXimum}
[SENSe:]PERiod:VOLTage:RANGe? [MINimum|MAXimum]

Function:
The commands set and query the voltage range under period measurement.




Advertising