Sense:]voltage:ac:range – RIGOL DM3058 User Manual

Page 125

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

4-32

The commands set and query the range of DC voltage measurement.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{200 mV|2 V|20 V

|200 V|1000 V}

20 V


Explanations:
[SENSe:]VOLTage:DC:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of DC voltage measurement from {200mV|2V|20V|200V|1000V}; thereinto,
MIN = 200 mV, MAX = 1000 V and the unit is V. If a 200 mV range is required,
enter 0.2 as its parameter.
[SENSe:]VOLTage:DC:RANGe? [MINimum|MAXimum] – Queries the range of DC
voltage measurement.

Return Format:
The query returns the range of DC voltage measurement.


3. [SENSe:]VOLTage:AC:RANGe

Command Format:
[SENSe:]VOLTage:AC:RANGe {<range>|MINimum|MAXimum}
[SENSe:]VOLTage:AC:RANGe? [MINimum|MAXimum]

Function:
The commands set and query the range of AC voltage measurement.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{ 200 mV|2 V|20 V

|200 V|750 V }

20 V


Explanations:
[SENSe:]VOLTage:AC:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of AC voltage measurement from {200 mV|2 V|20 V|200 V|750 V};
thereinto, MIN = 200 mV, MAX = 750 V and the unit is V. If a 200 mV range is
required, enter 0.2 as its parameter.