Sense:]resistance:nplc, Sense:]fresistance:nplc – RIGOL DM3058 User Manual

Page 140

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

4-47

26. [SENSe:]RESistance:NPLC

Command Format:
[SENSe:]RESistance:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum}
[SENSe:]RESistance:NPLC? [MINimum|MAXimum]

Function:
The commands set and query the integration time for 2-wire resistance
measurement. The inputted and returned values are multiple of PLC.

Explanations:
[SENSe:]RESistance:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum} – Sets the
desired integration time, MIN = 0.02 PLC, MAX = 100 PLC, and the default is 1
PLC.
[SENSe:]RESistance:NPLC? [MINimum|MAXimum] – Queries the current
integration time of 2-wire resistance measurement
The integration time relates to resolution and measurement rate, refer to Table
4-2.

Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
setting of resistance.

Return Format:
The query returns the current integration time of 2-wire resistance measurement.



27. [SENSe:]FRESistance:NPLC

Command Format:
[SENSe:]FRESistance:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum}
[SENSe:]FRESistance:NPLC? [MINimum|MAXimum]

Function:
The commands set and query the integration time for 4-wire resistance
measurement. The inputted and returned values are multiple of PLC.

Explanations:
[SENSe:]FRESistance:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum} – Sets the
desired integration time, MIN = 0.02 PLC, MAX = 100 PLC, and the default is 1