Calculate:rel:offset – RIGOL DM3058 User Manual

Page 31

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-6

Turn off the statistic function.
:CALCulate:STATistic:STATe OFF

7. :CALCulate:REL:OFFSet

Command Format:
:CALCulate:REL:OFFSet?
:CALCulate:REL:OFFSet {<range>|MIN|MAX|DEF|CURR}

Function:
The commands query and set the offset of REL operation.

Parameter:

Name

Type

Range

Default

<range>

Consecutive

real number

See Table 3-2

0


Explanations:
:CALCulate:REL:OFFSet? – Queries the REL offset of current measurement.
:CALCulate:REL:OFFSet {<range>|MIN|MAX|DEF|CURR} – Sets the desired REL
offset for current measurement.

The value of <range> varies with measurement type and range. See table below,
every parameter could be a value with sixth decimals. All the set value is only
valid for the current measurement and the value should be less than the upper
limit of current range.

Table 3-2 Offset range of REL operation

Measurement

items

Range

DEF

Unit

DC Voltage

±1200

0

V

AC Voltage

±900

0

V

DC Current

±12

0

A

AC Current

±12

0

A

Resistance

±1.2e+08

0

Ω

Capacitance

±1.2e-02

0

F

Frequency

±1.2e+06

0

Hz

Advertising
This manual is related to the following products: