Example 6: reading hold, Example 6: reading hold -13 – RIGOL DM3058 User Manual

Page 186

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

6-13

Example 6: Reading Hold


“Reading Hold” function can get a stable reading and hold it on the screen even when
test lead pens are removed.

The programming implementations——RIGOL Commands
0 *RST

/* Reset the instrument */

1 cmdset rigol /* Select RIGOL commands */
2 *IDN? /* Query the equipment ID and confirm if the multimeter works
normally */
3 /* Connect the test leads for DC voltage measurement */
4 :function:voltage:DC

/* Enable the DC voltage measurement */

5 :measure:voltage:DC <range> /* Set the measurement range */
6 :trigger:auto:hold on /* Enable the reading hold function */
7 :trigger:auto:hold:sensitivity 1 /* Set the holding range to “0.1%”*/


Explanations:
Command in line 7 permits multiple threshold ranges, for details, refer to Chapter 3.