Measure:period – RIGOL DM3058 User Manual

Page 65

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-40

Return Format:
The query returns: 0, 1, 2, 3 or 4 (see Table 3-5).

27. :MEASure:PERiod?

Command Format:
:MEASure:PERiod?

Function:
The command queries the measured period value.

Explanations:
The command is unavailable in double-display.

Return Format:
The command queries the measured period value of AC signal in scientific
notation, such as: 9.18543e-05, the unit is s.

28. :MEASure:PERiod

Command Format:
:MEASure:PERiod {<range>|MIN|MAX|DEF}

Function:
The command sets the voltage range under period measurement.

Parameter:

Name

Type

Range

Default

<range >

Discrete

{0|1|2|3|4}

2


Explanations:
The available period range is 1μs~50ms.
For the parameter range, refer to Table 3-5.

Example:
Sets the voltage range to 20V under period measurement.
:MEASure:PERiod DEF or :MEASure:PERiod 2

Advertising