Calculate:dbm:reference – RIGOL DM3058 User Manual

Page 99

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

4-6

Explanations:
The command is valid only when the dB function (CALCulate:FUNCtion DB) is
enabled under calculation is active (CALCulate:STATe ON).
CALCulate:DB:REFerence {<value>|MINimum|MAXimum} - Sets the desired dB
reference value.
CALCulate:DB:REFerence? [MINimum|MAXimum] - Queries the current dB
reference value.
MIN = -120 dB; MAX = 120 dB; DEF = 0 dB.

Return Format:
The query returns the dB reference value.



6. CALCulate:DBM:REFerence

Command Format:
CALCulate:DBM:REFerence {<value>|MINimum|MAXimum}
CALCulate:DBM:REFerence? [MINimum|MAXimum]

Function:
The commands set and query the dBm reference value.

Parameter:

Name

Type

Range

Default

<value>

Consecutive

Integer

2~8000

600


Explanations:
The command is valid only when the dB function (CALCulate:FUNCtion DBM) is
enabled under calculation is active (CALCulate:STATe ON).
CALCulate:DB:REFerence {<value>|MINimum|MAXimum} - Sets the desired
dBm reference value.
CALCulate:DB:REFerence? [MINimum|MAXimum] - Queries the current dBm
reference value.
MIN = 2Ω; MAX = 8000Ω; DEF = 600Ω.

Return Format:
The query returns the current dBm reference value.