Programming Guide for DM3058/DM3058E

5-20

1. COMP

Command Format:
COMP
COMP?

Function:
COMP – Enables the multimeter enter COMP mode, the Touch-Hold is enabled
automatically (or you can use HOLDCLR command to disable it).
COMP? - Queries the last comparison.

Return Format:
The query returns the last comparison. If the last COMP reading is higher than
the upper limit, the query returns HI; if it is less than the lower limit, returns LO;
if the result is within the range specified, returns PASS; if the measurement has
not been completed, returns “—”.



2. COMPCLR

Command Format:
COMPCLR

Function:
The command enables the multimeter exit the COMP mode and Touch Hold
function If it is active and backs into its normal work.



3. COMPHI

Command Format:
COMPHI <high value>

Function:
The command sets the upper limit of the COMP mode.

Explanations:
<high value> could be a signed integer or a signed real number.

Advertising