Configure:period – RIGOL DM3058 User Manual

Page 108

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

4-15

frequency measurement, but does not start the measurement.

Parameter:

Name

Type

Range

Default

<range>

Consecutive

Real Number

20 Hz~1 MHz

AUTO

<resolution>

Discrete

{MIN|MAX|DEF}

1ppm x range


Explanations:
All the units are Hz.
The <range> may be 20 Hz~1 MHz, thereinto, MAX = 1 000 000, MIN = 20, DEF
is auto range. Note that the multimeter uses one “range” for all inputs between
20Hz-1MHz.



8. CONFigure:PERiod

Command Format:
CONFigure:PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}

Function:
The command presets the multimeter as specified range and resolution for period
measurement, but does not start the measurement.

Parameter:

Name

Type

Range

Default

<range>

Consecutive

Real Number

1 μs~50 ms

50 ms

<resolution>

Discrete

{MIN|MAX|DEF}

1ppm x range


Explanations:
The units is s.
The <range> may be 1 μs~50 ms, thereinto, MAX = 0.05, MIN = 0.000001, DEF
= 0.05. Note that the multimeter uses one “range” for all inputs between
1μs-50ms.