Configure:period – RIGOL DM3058 User Manual
Page 108
RIGOL
Programming Guide for DM3058/DM3058E
4-15
frequency measurement, but does not start the measurement.
Parameter:
Name
Type
Range
Default
<range>
Consecutive
Real Number
20 Hz~1 MHz
AUTO
<resolution>
Discrete
{MIN|MAX|DEF}
1ppm x range
Explanations:
All the units are Hz.
The <range> may be 20 Hz~1 MHz, thereinto, MAX = 1 000 000, MIN = 20, DEF
is auto range. Note that the multimeter uses one “range” for all inputs between
20Hz-1MHz.
8. CONFigure:PERiod
Command Format:
CONFigure:PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Function:
The command presets the multimeter as specified range and resolution for period
measurement, but does not start the measurement.
Parameter:
Name
Type
Range
Default
<range>
Consecutive
Real Number
1 μs~50 ms
50 ms
<resolution>
Discrete
{MIN|MAX|DEF}
1ppm x range
Explanations:
The units is s.
The <range> may be 1 μs~50 ms, thereinto, MAX = 0.05, MIN = 0.000001, DEF
= 0.05. Note that the multimeter uses one “range” for all inputs between
1μs-50ms.