RIGOL DM3058 User Manual

Page 68

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-43

Return Format:
The query returns the measured capacitance value in scientific notation, such as:
8.889030e-05, the unit is F.



34. :MEASure:CAPacitance

Command Format:
:MEASure:CAPacitance {<range>|MIN|MAX|DEF}

Function:
The command sets the capacitance range.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{0|1|2|3|4|5}

2

Explanations:
For the parameter range, see table below.

Table 3-9 Capacitance range

Parameter

Range

0

2 nF

1

20 nF

2

200 nF

3

2 μF

4

200 μF

5

10000 μF

MIN

2 nF

MAX

10000 μF

DEF

200 nF

Example:
Sets the capacitance range to 2 μF.
:MEASure:CAPacitance 3

35. :MEASure:CAPacitance:RANGe?

Command Format:

Advertising