Programming Guide for DM3058/DM3058E

4-33

[SENSe:]VOLTage:AC:RANGe? [MINimum|MAXimum] – Queries the range of AC
voltage measurement.

Return Format:
The query returns the range of AC voltage measurement.



4. [SENSe:]CURRent:DC:RANGe

Command Format:
[SENSe:]CURRent:DC:RANGe {<range>|MINimum|MAXimum}
[SENSe:]CURRent:DC:RANGe? [MINimum|MAXimum]

Function:
The commands set and query the range of DC current measurement

Parameter:

Name

Type

Range

Default

<range>

Discrete

{ 200 µA|2 mA|20 mA

|200 mA|2 A|10 A }

200 mA


Explanations:
[SENSe:]CURRent:DC:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of DC current measurement from {200 µA|2 mA|20 mA|200 mA|1 A|10 A};
thereinto, MIN = 200 µA, MAX = 10 A and the unit is A. If a 200 mA range is
required, enter 0.2 as its parameter.
[SENSe:]CURRent:DC:RANGe? [MINimum|MAXimum] – Queries the range of DC
current measurement.

Return Format:
The query returns the range of DC current measurement.



5. [SENSe:]CURRent:AC:RANGe

Command Format:
[SENSe:]CURRent:AC:RANGe {<range>|MINimum|MAXimum}
[SENSe:]CURRent:AC:RANGe? [MINimum|MAXimum]