Relclr, Relset – RIGOL DM3058 User Manual

Page 162

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

5-13

13. REL

Command Format:
REL

Function:
The command enables the multimeter enter the REL mode, using the value
shown on the main display as the relative base.



14. RELCLR

Command Format:
RELCLR

Function:
The command enables the multimeter exit the REL mode and backs into ranging
mode. The range selected is the one prior to REL.



15. RELSET

Command Format:
RELSET <relative base>
RELSET?

Function:
RELSET <relative base> - Enables the multimeter enter the REL mode, using the
<relative base> as the relative base.
RELSET? – Queries the current relative base. If the REL operation is disabled, an
Execute error is generated.

Explanations:
<relative base> could be a signed integer, a signed real number with exponent or
a signed real number without exponent. If <numeric value> is overrange, an
Execution Error may generate.

Return Format:
The query returns the relative base of REL operation.

Advertising