RIGOL DM3058 User Manual

Page 13

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

2-4

1. *CLS

Command Format:
*CLS

Function:
The command resets the values in all Event Registers and clears the Error Queue
information.



2. *ESE

Command Format:
*ESE <enable_value>
*ESE?

Function:
The commands set and query the enable value of Event Status Register (ESR).

Explanations:
*ESE <enable_value> - Sets the desired enable value of ESR.
*ESE? - Queries the current enable value of ESR.
<enable_value> - Ranging from 0 to 189.

Return Format:
The query returns the enable value of ESR.



3. *ESR?

Command Format:
*ESR?

Function:
The command queries the event value of Event Status Register (ESR).

Return Format:
The query returns the current event value of ESR.