Measure:voltage:dc – RIGOL DM3058 User Manual

Page 53

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-28

The command returns the measured DC voltage value in scientific notation, such
as: 8.492853e-05, the unit is V.


4. :MEASure:VOLTage:DC

Command Format:
:MEASure:VOLTage:DC {<range>|MIN|MAX|DEF}

Function:
The command sets the range and resolution of DC voltage measurement.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{0|1|2|3|4}

2


Explanations:
The corresponding resultion will be set automatically after you set the range.
The measurement type will change into “Manual” automaticlly as you set the
range.
For the parameter range and the corresponding resolution, see table below.

Table 3-4 DC voltage range and

the

corresponding resolution

Parameter

Range

Resolution

0

200 mV

100 nV

1

2 V

1 μV

2

20 V

10 μV

3

200 V

100 μV

4

1000 V

1 mV

MIN

200 mV

100 nV

MAX

1000 V

1 mV

DEF

20 V

10 μV