Calculate:db:reference – RIGOL DM3058 User Manual

Page 98

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

4-5

CALCulate:LIMit:UPPer {<value>|MINimum|MAXimum}
CALCulate:LIMit:UPPer? [MINimum|MAXimum]

Function:
The commands set and query the upper limit of current measurement.

Parameter:

Name

Type

Range

Default

<value>

Consecutive

Integer

±120% of MAX range

0


Explanations:
The command is valid only when the LIMIT function (CALCulate:FUNCtion LIMIT)
is enabled under calculation is active (CALCulate:STATe ON).
CALCulate:LIMit:UPPer {<value>|MINimum|MAXimum} - Sets the desired upper
limit under current measurement.
CALCulate:LIMit:UPPer? [MINimum|MAXimum] - Queries the upper limit under
current measurement.
MIN denotes the minimum range (-120% of MAX range) of current measurement.
MAX denotes the maximum range (120% of MAX range) of current measurement.

Return Format:
The query returns the upper limit of current measurement.



5. CALCulate:DB:REFerence

Command Format:
CALCulate:DB:REFerence {<value>|MINimum|MAXimum}
CALCulate:DB:REFerence? [MINimum|MAXimum]

Function:
The command commands set and query the dB reference value.

Parameter:

Name

Type

Range

Default

<value>

Consecutive Integer

-120 dB~+120 dB

0

Advertising