Trigger:single – RIGOL DM3058 User Manual

Page 84

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-59

Explanations:
:TRIGger:AUTO:HOLD:SENSitivity? – Queries the current delay sensitive of auto
trigger.
:TRIGger:AUTO:HOLD:SENSitivity {<value>|MIN|MAX|DEF} Sets the desired
delay sensitive of auto trigger. For the parameter range, see table below:

Table 3-11 Delay sensitive of auto trigger

Parameter

Sensitive

0 (MIN)

0.01%

1

0.1%

2 (DEF)

1%

3 (MAX)

10%


Return Format:
The query returns: 0, 1, 2 or 3 (see table 3-11).



5. :TRIGger:SINGle

Command Format:
:TRIGger:SINGle?
:TRIGger:SINGle {<value>|MIN|MAX|DEF}

Function:
The commands query and set the number of sample in single trigger mode.

Explanations:
:TRIGger:SINGle? – Queries the number of sample under current single trigger.
:TRIGger:SINGle {<value>|MIN|MAX|DEF} – Sets the required sample number
in single trigger mode.

Parameter:

Name

Type

Range

Default

<value>

Consecutive

Integer

1~2000

1


Return Format:
The query returns the number of sample under current single trigger.