Sense:]fresistance:resolution, Sense:]voltage:dc:nplc – RIGOL DM3058 User Manual

Page 138

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

4-45

setting of resistance.

Return Format:
The query returns the current precision of 2-wire resistance measurement.



23. [SENSe:]FRESistance:RESolution

Command Format:
[SENSe:]FRESistance:RESolution {<resolution>|MINimum|MAXimum}
[SENSe:]FRESistance:RESolution? [MINimum|MAXimum]

Function:
The commands set and query the precision of 4-wire resistance measurement.

Parameter:

Name

Type

Range

Default

<resolution> Discrete

Refer to Table 4-2

3ppm x range


Explanations:
[SENSe:]FRESistance:RESolution {<resolution>|MINimum|MAXimum}– Sets the
desired precision of 4-wire resistance measurement
[SENSe:]FRESistance:RESolution? [MINimum|MAXimum] – Queries the current
precision of 4-wire resistance measurement.
<resolution> relates to measurement rate and integration time, refer to Table
4-2.

Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
setting of resistance.

Return Format:

The query returns current precision of 4-wire resistance measurement.



24. [SENSe:]VOLTage:DC:NPLC

Command Format:
[SENSe:]VOLTage:DC:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum}
[SENSe:]VOLTage:DC:NPLC? [MINimum|MAXimum]