Measure:fresistance:range, Measure:frequency – RIGOL DM3058 User Manual

Page 63

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-38

:MEASure:FRESistance {<range>|MIN|MAX|DEF}

Function:
The command sets the desired 4-wire resistance range.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{0|1|2|3|4|5|6 }

3


Explanations:

For the parameter range, refer to Table 3-8.

Example:
Sets the 4-wire resistance range to 200Ω.
:MEASure:FRESistance MIN or :MEASure:FRESistance 0



23. :MEASure:FRESistance:RANGe?

Command Format:
:MEASure:FRESistance:RANGe?

Function:
The command queries the current 4-wire resistance range.

Return Format:
The query returns: 0, 1, 2, 3, 4, 5 or 6 (see Table 3-8).

24. :MEASure:FREQuency?

Command Format:
:MEASure:FREQuency?

Function:
The command queries the measured frequency value.
Explanations:
The available frequency range is 20Hz~1MHz.
The command is unavailable in double-display.