RIGOL DM3058 User Manual

Page 16

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

2-7

*STB?

Function:
The command queries the condition value of Status Byte Register (STB).

Return Format:
The query returns the current condition value of STB.

10. *TRG

Command Format:
*TRG

Function:
The command generates a trigger signal when the instrument is in
“Wait-for-trigger” state.



11. *TST?

Command Format:
*TST?

Function:
The command queries the self-test result.

Return Format:
The query returns 0 or 1. Thereinto, 0 denotes pass, and 1 denotes failure.



12. *WAI

Command Format:
*WAI

Function:
The command waits until all pending operations complete.