Calculate:pf:lower – RIGOL DM3058 User Manual

Page 36

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-11

Explanations:
When the measurement result is within the maximum and minimum, the PF
operation is success.

Return Format:
The query returns the PF operation: PASS, HI or LO.



16. :CALCulate:PF:LOWEr

Command Format:
:CALCulate:PF:LOWEr?
:CALCulate:PF:LOWEr {<range>|MIN|MAX|DEF}

Function:
The comands query and set the lower limit of P/F operation.

Parameter:

Name

Type

Range

Default

<range>

Consecutive

real numbe

See Table 3-3

0


Explanations:
:CALCulate:PF:LOWEr? – Queries the lower limit of P/F operation
:CALCulate:PF:LOWEr {<range>|MIN|MAX|DEF} – Sets the lower limit of P/F
operation. See table below for more details.

Table 3-3 Ranges of P/F operation

Measurent

items

Range

Unit

DC Voltage

±1200

V

AC Voltage

0~900

V

DC Current

±12

A

AC Current

0~12

A

Resistance

0~1.2e+08

Ω

Capacitance

0~1.2e-02

F

Period

1.0e-06~100

s

Frequency

0~1.2e+06

Hz

Advertising
This manual is related to the following products: