Measure:resistance – RIGOL DM3058 User Manual

Page 61

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-36

Command Format:
:MEASure:RESistance?

Function:
The command queries the measured 2-wire resistance value.

Explanations:
The command is unavailable in double-display.

Return Format:
The command queries the measured 2-wire resistance value in scientific notation,
such as: 8.366031e-05, the unit is Ω.



19. :MEASure:RESistance

Command Format:
:MEASure:RESistance {<range>|MIN|MAX|DEF}

Function:
The command sets the desired 2-wire resistance range.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{0|1|2|3|4|5|6}

3


Explanations:
For the parameter range, see table below.

Table 3-8 2-wire resistance range

Parameter

Range

0

200 Ω

1

2 kΩ

2

20 kΩ

3

200 kΩ

4

1 MΩ

5

10 MΩ

6

100 MΩ