Trigger:source, Trigger:auto:interval – RIGOL DM3058 User Manual

Page 82

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

3-57

1. :TRIGger:SOURce

Command Format:
:TRIGger:SOURce?
:TRIGger:SOURce {AUTO|SINGLE|EXT}

Function:
The commands query and set the trigger source.

Explanations:
:TRIGger:SOURce? – Queries the trigger source used currently.
:TRIGger:SOURce {AUTO|SINGLE|EXT} – Sets the trigger source to be used. The
parameter AUTO, SINGLE, EXT denotes Auto trigger, Single trigger and External
trigger, separately.

Return Format:
The query returns: AUTO, SINGLE or EXT.



2. :TRIGger:AUTO:INTErval

Command Format:
:TRIGger:AUTO:INTErval?
:TRIGger:AUTO:INTErval <value>

Function:
The commands query and set the interval of auto trigger.

Parameter:

Name

Type

Range

<value>

Consecutive

Integer

See Explanations below


Explanations:
:TRIGger:AUTO:INTErval? – Queries the interval of auto trigger.
:TRIGger:AUTO:INTErval <value> – Sets the desired interval of auto trigger.
Different rate has different intervals:
In Fast measruement, the default interval is 8 ms and the available range is within
8 ms and 2000 ms;

Advertising