FEC AFC1500 User Manual

Page 158

Advertising
background image

Chapter 9: Troubleshooting

Page 9-4

9.2.4 Code 1-3 TORQUE TRANSDUCER / CAL SELF CHECK ERROR

Calibration level voltage error after a fastening start was attempted.

This abnormal is caused from;
1. When the tool transducer is sensing excessive torque due to pressure or vibration during
the fastening cycle start.
2. If the controller or the transducer cable is located in an electric or magnetic noise field.
3. When the torque transducer, transducer (T/D) cable or the controller malfunctions.

RECOVERY:
1. Check that the tool has no external force applied to it.
2. Verify the tool mounting bolts are tight and nothing is cause excessive vibration to the tool.
3. Verify that the cables or controllers are not located near any high voltage transient power
sources. Relocate as required and attempt again a fastening start input.
4. Exchange transducer (T/D) cable, tool and/or controller with known working units. Input the
start input after each exchange, and make note of any change in the location of the abnor-
mal. 5. Replace the defective component.

9.2.5

Code 1-4 TORQUE TRANSDUCER/STARTED ON ZERO ERROR CONDITION

The start signal was input while a Zero voltage abnormal condition existed.

RECOVERY:
1. Interlock the PLC circuit to disable the start signal during an existing abnormal condition.
2. Follow the ZERO VOLTAGE ERROR abnormal help procedure.

9.2.6 Code 1-5

TORQUE TRANSDUCER / STARTED ON CAL ERROR CONDITION

The start signal was input while a Cal voltage abnormal condition existed.

RECOVERY:
1. Interlock the PLC circuit to disable the start signal during an existing abnormal condition.
2. Follow the CAL VOLTAGE ERROR abnormal help procedure.

9.2.7 Code 1-6

TORQUE TRANSDUCER / ZERO LEVEL SELF CHECK ERROR

Zero level voltage loaded to memory from tool EEPROM during initialization does not match
the actual zero level during a self check with the self check function enabled.

This abnormal is caused from;
1. When the tool transducer is sensing excessive torque due to pressure on the tool body.
2. If the controller or the transducer cable is located in an electric or magnetic noise field.
3. When the torque transducer, transducer (T/D) cable or the controller malfunctions.

RECOVERY:
1. Check that the tool has no external force applied to it.
2. Verify that the cables or controllers are not located near any high voltage transient power
sources. Relocate as required and reinitialize.
3. Exchange transducer (T/D) cable, tool and/or controller with known working units. Reini-
tialize the system after each exchange, and make note of any change in the location of the
abnormal.
4. Replace the defective component.

Advertising