1 wiring precautions – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 100

Advertising
background image

IM 11M12D01-01E

6-2

6.1.2

High-Temperature Detector (Model ZR22G-015)

Sample gas outlet
When a negative measurement gas
pressure is used, connect the auxiliary
ejector assembly.
When the measurement gas is
high-temperature and high-pressure,
and does not fall below 700

Њ

C,

connect a pressure control valve
(e.g. a needle valve).
(Refer to Section 3.2.2.)

Separate type
High-temperature Detector
(ZR22G-015)
When the temperature
of the measurement
gas is between 700

Њ

and 1400

Њ

C,

mount this detector
with a ZO21P-H probe adapter.

F0602e.eps

Flange
Selectable from JIS standards
or ANSI standards

High-temperature Probe Adapter
(ZO21P-H)
The probe is made of either SUS 310S
or silicon carbide (SiC).
Its length is either 1.0 m or 1.5 m.
When using an SiC probe,
mount it vertically downward.

Figure 6.2 High-temperature Detector