Caution – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 55

Advertising
background image

IM 11M12D01-01E

3-5

3. Installation

3.1.7 Detector with a probe protector (Model ZO21R-L-

ᮀᮀᮀ- *B for dust wear

protect

The detector is used with a probe protector to prevent the sensor from being worn by
dust particles when there is a high concentration of dust and gas flow exceeds 10m/s
(fine-carbon boiler or fluid-bed furnace).

(1) Put a gasket (provided by the user) between the flanges, and mount the probe

protector in the probe insertion hole. The probe protector should be installed so that
the notch is downstream of the measurement gas flow.

(2) Make sure that the cell assembly mounting screws (four) at the probe tip are not

loose.

(3) Where the detector is mounted horizontally, the reference gas and calibration gas

inlet should face downward.

CAUTION

When the probe protector is used in the Zirconia Oxygen Analyzer version with
pressure compensation (-P), instrument air leaking from the probe protector may
affect the measured value.

Direction of the measurement gas flow

Mount the protector so that the notch is
or the downstream side of the gas flow.

Reference gas inlet
Calibration gas inlet

Unit: mm

F3.4E.EPS

Probe top

1050,1550,2050

Gasket (t1.5)

Figure 3.4 Mounting of detector with a probe protector