3.3.1

Installation Location

Follow the guidelines below when selecting a location for the averaging converter.

(1) The touch screen display should be clearly visible and easily accessible for key

operation: the display at eye level is recommended.

(2) A wall or construction should be strong enough to support the weight of the

converter.

(3) The converter should be easily accessible for inspections and maintenance.
(4) The ambient temperature should be in the range of -5 to 50

ЊC and the temperature

variation should be small: the daily temperature range within 15

ЊC is recommended.

(5) The humidity should be in the range of 10 to 85%RH and no condensation may

occur.

(6) No corrosive gases are present.
(7) No dust exists.
(8) There is minimal vibration.
(9) The converter should be near the detectors: the two-way resistance between detector

and converter should not exceed 10

⍀.

DANGER

• The averaging converter is heavy. Take extra care not to accidentally drop it.

Installation work should be done by an adequate number of people.

CAUTION

The averaging converter is heavy (100/115 VAC version: approx. 13 kg, 230 VAC

version: approx. 25 kg) so needs to be installed on a wall or construction strong
enough to support.

3.3.2

Installation in an Instrument Panel

When installing the averaging converter in an instrument panel, operability,
maintainability, and environmental resistance should be considered.

(1) Installation Space and Relationship to Peripheral Devices

Follow the instructions below to determine the positions of the averaging converter and
peripheral devices.

• Do not place heat-generating devices in close proximity of the averaging converter:

specially, avoid a place under the averaging converter.

• Do not place power related devices such as transformers or electromagnetic switches

in close proximity of the averaging converter.

• Segregate wiring ducts for signal lines and those for power lines and do not place in

close proximity.

Figure 3.8 shows the requirements for the mounting holes for the averaging converter.
See Section 2.4 for the dimensions of the averaging converter and maintenance
clearance.