7.7 Current Output Range Setting

The output ranges (two ranges) of each channel are independently settable. Ranges are
switched by contact input. Contact OFF results in Range 1, and contact ON results in
range 2. For details, refer to Sec. 8.5 Setting Contact Inputs. Shipping time default is
for each range to be set to 0 to 25% O

2

. The ratio of the maximum range setting to the

minimum is restricted as described in Sec. 8.1 Setting Analog Outputs.

1) Touch the Setup key to switch to the [Execution/Setup] display.
2) From the [Commissioning] display select [mA-output setup], then [Set Range]. (See

Fig. 7.6).

3) Use [

᭣ ] and [᭤] keys to select the channel.

4) Set [Min.] O

2

concentration corresponding to 4 mA output, then set [Max.] O

2

concentration corresponding to 20 mA output (select, then touch Enter key).

5) The numeric data entry display shown in Fig. 7.7 appears; enter the desired oxygen

concentration setting. To set 10% O

2

, input [010] and touch Enter.

6) As necessary, repeat for each averaging output a, b and c of other channels.

Enter

mA-output range

max: 25 %O2

Range1

Range2

F7.6E.EPS

min: 0 %O2

max: 25 %O2

min: 0 %O2

Ch1

Enter

New value:

0

1

4

7

2

3

5

6

8

9

025 %O2

F17E.EPS

Fig. 7.6 Current-output range Fig. 7.7 Numerical entry