5 other settings, Others, Clock – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 143: 1 date and time

Advertising
background image

IM 11M12D01-01E

8-17

8. Detailed Data Setting

8.5 Other Settings

8.5.1 Date and Time

Here we explain how to set date and time. Date and time setting are used for Auto
Calibration and Auto Indication Check, also Auto Blowback Start time.

<Setting Procedure>
1) From the basic panel display if you touch the Setup key, the [Execution/Setup]

display appears.

2) Select [Commissioning] and the Commissioning display appears.
3) Select [Others] and the display of Fig 8.17 appears.
4) Select [Clock] and the display of Fig 8.18 appears.
5) Select [Set date] and the numerical entry display appears. To enter June 21. 2004 ,

enter [210604]. Touch the Enter key and you revert to the display of Fig 8.17.

6) Select [Set time]. Enter the time in 24-hour format. To set 2:30 pm, enter [1430] on

the numerical display and touch enter. Touch the Enter key. The clock starts from
00 seconds.

Enter

F8.17E.EPS

Clock

Averaging

Fuel setup

Purging

Unit setup

Asymmetry alarm

Passwords

Defaults

Others

Fig. 8.17 Others setting display

Enter

F47E.EPS

Set date: 2 1 / 0 6 / 0 4

Set time: 1 6 : 4 4

2 1 J un. 2 0 0 4

1 6 : 4 4 : 4 5

Clock

Fig. 8.18. Day and Time setting

Advertising