Manual cal – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 121

Advertising
background image

IM 11M12D01-01E

7-11

7. Startup

8) Open the span gas flow setting valve, and adjust flow to 600

Ϯ60 ml/min.

9) Select [Valve opened (Start Calibration)] and touch the [Enter] key. An oxygen

concentration graph like Fig. 7.13 is displayed, and [CAL.TIME] status mark flashes.
Wait for the graph measured value trace to stabilize. Calibration has not yet been
performed, so it is not a problem if the displayed value is different from the correct
value of span gas concentration.

F7.13E.EPS

Enter

21.00%

Ch1

1min./div.

0.98%

21.0 %O2

CAL. TIME

Manual cal.

Channel No.

under calibration

Setting value of

span gas concentration

Setting value of

zero gas concentration

Actual measurement value

Fig 7.13 Display during calibration

10) When the graph trace stabilizes, touch the [Enter] key. The display changes to like

Fig. 7.14. The measured value is adjusted to read the same as the span gas concentra-
tion set value, and a check mark appears in the check box beside Span calibration.
Close the span gas flow setting valve and tighten the lock nut used to prevent it from
loosening during measurement.

11) Next perform zero calibration. Select [Zero calibration] and touch the [Enter] key.
12) Select [Next] and touch the [Enter] key.

Enter

F7.14E.EPS

Ch1

Manual cal.

Span calibration

Zero calibration

Zero calibration

End

Close the span gas valve.

Fig. 7.14 Zero calibration start display

13) Open the zero gas flow setting valve, and adjust flow to 600

Ϯ60 ml/min.