Manual ind. check, 5 running an indication check – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 177

Advertising
background image

IM 11M12D01-01E

10-15

10. Other Functions

10.4.5 Running an Indication Check

10.4.5.1 Manual Indication Check

A manual indication check is performed by following prompts on the screen, and
manually opening or closing valves of calibration gas or third check gas. When solenoid
valves are used, valves are opened and closed automatically at the same time that
messages appear on the screen. The procedure for indication check is virtually the same
as for calibration. Taking a 3-point indication check using zero, span, and third gases for
example, the procedure is described below.

<Procedure>
1) Confirm that zero, span, and third gas flow control valves are fully closed, then adjust

supply gas pressure to a sample gas pressure + 50 kPa (but a maximum of 300 kPa)
using regulator valves or air sets on the zero, span and third check gas cylinders.

2) Touch [Setup] on the basic panel display to switch to the Execution/Setup display.
3) Touch [Indication check] and the Indication Check display appears.
4) Touch [Manual Ind. check] then the Enter key. The Manual Ind. check Channel

Select display shown in Figure 10.13 appears.

Enter

Manual ind. check

Ch1

Ch2

Ch3

Ch4

Ch5

Ch6

Ch7

Ch8

F10.13E.EPS

Select ind. check channel

Enter

Zero gas ind. check

Third gas ind. check

F10.14E.EPS

Manual ind. check

Span gas ind. check

Ch2

Fig. 10.13 Channel select screen Fig. 10.14 Manual Span Ind. check

5) Select the channels to be checked (channels in warm-up or error status cannot be

selected). Here Channel 2 is taken as an example.

6) Select [Span indication check] and touch the Enter key. A display like Figure 10.15

appears.

7) Open the span gas flow control valve, and adjust flow rate to 600

Ϯ60 ml/min. If

auto indication check piping and wiring are installed, then span gas starts to flow
automatically.

8) Touch [Start ind. check] touch the Enter key (Figure 10.15). An oxygen concentration

trend graph like Figure 10.16 is displayed, and a [CHK.TIME] status mark blinks.
Watch the trend graph and confirm that the measured value stabilizes. Compare the
measure span gas concentration displayed at the bottom of the display against the
certified concentration of span gas to check for any discrepancy or error.

Enter

F10.15E.EPS

Ch2

Manual ind. check

Open span gas valve.
Set flow span gas to
600ml/min.

Start ind. check

Cancel ind. check

Fig 10.15 Manual Indication Check Start