3 piping for the reference gas, 4 piping to the high temperature probe adapter, 2 wiring holes – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 84

Advertising
background image

IM 11M12D01-01E

5-4

5.1.2

Wiring Holes

The averaging converter has 30 wiring holes. Use holes near the rear for power wiring
(e.g., power cables, detector heater wires, and wiring for solenoid valve contacts and
other contact outputs). For signal wiring (e.g., detector signal lines and analog output
signal lines), use holes near the front.

If more than 30 cables are used and thus more than one cable has to be passed through
one wiring hole, do not allow signal lines and power lines to pass the same hole. Also,
in the averaging converter, the signal lines and power lines should be routed so they can
avoid contact and interference.

A grommet is attached, as standard, to the wiring hole of the averaging converter. An
optional cable gland is available. When a cable gland is not used, make a cable hole in a
grommet.

Wiring Connections:

30 holes

Applicable Cable Hole:

л17 mm maximum for grommet
л6 to л12 mm for cable gland (optional)

5.1.3

External Wiring Connection Terminals of the Averaging Converter

The connection diagram of the averaging converter is shown in Figure 5.3. The terminal
numbers are indicated on the converter. Care should be taken to make the wiring
connections correctly.

+ 1

+ 1

+

1

+

1

+ 1

- 2

- 2

-

2

-

2

- 2

+ 3

+ 3

+

3

+

3

+ 3

- 4

- 4

-

4

-

4

- 4

+ 5

+ 5

+

5

+

5

+ 5

- 6

- 6

-

6

-

6

- 6

DI1 7

+ 7

+

7

+

7

+

7

DI2 8

- 8

-

8

-

8

- 8

COM 9

C

NC NO

C

NC NO

C NC NO

C NC NO C

NC NO

C

NC NO

C

NC NO

11

12

13

14

15

16

17

19

20 21

22

24

25

26

27

28 29

61 62

63

64

65

66

68

70 71 72

31 32

33

34

35

36

37

39

40 41

42

44

45

46

47

48 49

81 82

83

84

85

86

88

90 91 92

L

C

NC NO

C

NC NO

C NC NO

C NC NO

C

NC NO

C

NC NO

N

G

51 52 53 54

55 56

57 58

101 102 103 104

108

1

2 3

4

5

6

2

4 5 6

7

8

7 8

SV-CH1 SV-CH2

DO1

DO2

DO3

DO4

DO5

SV-CH6

SV-CH5

CELL

TC

CJ

TC

CJ

AO

TC

CJ

AO

TC

CJ

AO

Ave-a

Ave-b

Ave-c

CELL

CELL

CELL

CH1 CH5

SV-COM SV-CH3 SV-CH4

DO-CH1

DO-CH2

DO-CH3

DO-CH4

CH1

CH2

CH3

CH4

CH5

CH6

CH7

CH8

DO-CH7

DO-CH8

DO-CH5

DO-CH6

SV-CH8

SV-CH7

105 106 107

1

3

87

89

67

69

23

43

18

38

Detector

Detector

M3.5 screws

M4 screws

Control Card

Channel Card (CH1)

Channel Card (CH4)

Channel Card (CH5)

Channel Card (CH8)

Detector
Output
signals

Detector
Output
signals

Detector
Output
signals

Detector
Output
signals

Individual
Analog
output

Individual
Analog
output

Individual
Analog
output

Individual
Analog
output

AO

Averaged
analog outputs

Contact input

Functional Contact Output

Common Error output Solenoid Valve output

Individual Error Contact Output

Solenoid Valve Contact Output

Basic Power Supply

Expansion Power Supply

Solenoid Valve Contact Output

Solenoid Valve Contact Output

Individual Error Contact Output

Individual Error Contact Output

Cell Heater

Cell Heater

Grounding plate

Power
supply

Ground

External
ground
terminal

G terminal and

terminal are connected by jumper plate.

Power supply and Ground terminals are M4 screw.

If 230 V AC is selected, channel cards are expandable up to 4.

ء

2

ء

1 Averged analog output with digital communication (HART) when suffix

code “-E” (HART communication) is selected.

ء

1

ء

2 Used exclusively for communication when suffix code “-F” (FOUNDATION

Fieldbus communication) is selected.

Figure 5.3 External Wiring Connection Terminals of the Averaging Converter

Advertising