9. Calibration

3) Calibration time setting

If the mode is [Man]:
Select [Hold time]. This is the output stabilization (settling) time required from calibra-
tion end until measurement mode is reentered. During this time, the calibration gas is
switched to measurement gas and the measurement is allowed time to stabilize. Setting
range is from 00min. 00 sec. to 60 min 59 sec. The analog outputs of individual chan-
nels which are under calibration remain in Output Hold status from the start of calibra-
tion until the end of the stabilization (settling) time, and their output is not included in
averaging group calculations during this time.

If the mode is [Semi-Auto]:
In addition to the above [Hold] output stabilization (settling) time, there is also a [Cal.
time] setting. This is the settling time from when calibration gas flow is started until
calibration is performed. The same time settingis used for both Zero and Span calibra-
tion. Setting range is from 00min. 00 sec. to 60 min 59 sec. Calibration time and
stabilization time, and analog output status are shown on the calibration timing chart of
Fig. 9.6 below. For Solenoid Valve (SV) piping and wiring see Sec. 4.2 (on
AutoCalibration).

Ch1 Span CAL

CAL time

Stabilizn.

CAL start

Off

On

On

Off

Zero gas ON

Span gas on

Zero gas ON

Span gas ON

Cal start in

F9.6E.eps

Ch1 Zero CAL

Ch1 output hold

Ch2 Span CAL

Ch2 Zero CAL

Ch2 output hold

Average output

SV-Ch1 contact

SV-Ch2 contact

SV-Ch2 contact

Ch1 removed from average

Ch2 removed from average

Ch.2 Output hold

CAL time

Ch.1 Output hold

CAL time

Stabilizn.

CAL time

Fig. 9.6 Calibration Timing