IM 11M12D01-01E

5-7

5. Wiring

The sizes of the terminal screws are M3.5. Each wire should be terminated with a crimp
terminal appropriate to the screw.

WARNING

If the ambient temperature at the detector installation site exceeds 60

ЊC, use bare

crimp terminals.

• Before opening the detector cover, loosen the lock screw. If the screw is not loosened

first, the screw will damage the cover, and the terminal box will require replacement.
When opening and closing the cover, remove any sand particles or dust to avoid
gouging the thread.

• Notice when closing the cover of the detector

After screwing the cover in the detector body, secure it with the lock screw.

Lock Screw

Detector Cover

F5.8E.EPS

Figure 5.5