7 cell response time, 8 robustness of a cell, 9 cell’s internal resistance – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 197: Caution, Channel card power, Danger

Advertising
background image

IM 11M12D01-01E

11-3

11.

Inspection and Maintenance

Channel Card

CELL, TC, CJ

F11.2E.eps

Detector

Power supply wire for Heater

Averaging Converter

Power Supply Circuit

Hot Swap Switch Mechanical Relays

Fig. 11.2

<Procedure>

1) On the basic panel display touch the Setup key. The Execution/Setup display will

appear.

2) Use the Cursor key to move the cursor to Maintenance. Touch the Enter key.
3) On the Maintenance display, select Channel Card Power. Touch the Enter key. A

display shown in Figure 11.3 will appear.

4) On the Channel Card Power display, select the channel of a detector to be discon-

nected for inspection or maintenance. Touch the Enter key.

5) A dropdown menu (Disable/Enable) will appear. Select Disable and touch the Enter

key. Make sure that all status lamp(s) of disabled channel(s) have turned off.

Enter

Ch1 : ZR22G

Ch2 : Enable

Ch3 : Enable

Ch4 : Enable

Ch5 : Enable

Ch6 : Enable

Ch7 : Enable

Ch8 : Enable

F11.3E.EPS

Disable

Channel card power

Figure 11.3 Channel Card Power (Hot Swap) Display

DANGER

Before the inspection or maintenance of a detector, make sure that the power has been
turned off by measuring the voltage of the heater terminal in the terminal box of the
detector for more than two seconds with a voltmeter.

CAUTION

The state (Enable/Disable) of the hot swap function is stored even after the power switch
of the averaging converter is turned off. When the power switch of the averaging
converter is turned on, power is not supplied to the disabled channel(s).