7.8 Averaging Group Setting

This instrument can determine oxygen concentration average for three averaging groups
a, b, and c. a and b can each be arbitrary channel groupings, and the corresponding
averages are output. c output is (a+b) /2. Select the groupings for a and b as per the
procedure below.

1) Touch the Setup key to switch to the [Execution/Setup] display.
2) From the [Commissioning] display select [Average group setup], then [Average group

-a]. Fig. 7.8 shows the display.

3) Select the group to set (a or b) then touch Enter. Channels that you can select are

displayed.

4) Channels that are set to [ON] belong to that group. Use the pointer cursor keys to

move the pointer to any channel whose status you want to toggle, then touch the
Enter key. The ON/OFF selection window opens, and you can toggle status.

5) Any or all channels may belong to both groups a and b.

Enter

Average group-a

Ch1 ON

Ch2 ON

Ch3 ON

Ch4 ON

Ch5 OFF

Ch6 OFF

Ch7 OFF

Ch8 OFF

F7.8E.EPS

Fig. 7.8 Setting Averaging Group

< Shipping-time defaults >
At shipping time, averaging groups a and b contain all channels (all channels [ON).

Note

If you set all channels in a group to [OFF] then the averaged output will be held at

the current set by [Ave group invalid] on the [mA-outputs hold] display. In this case,
the status mark at the bottom of the display will be [HOLD]. For details, refer to Sec.
8.1.2 Output Hold Settings.