2 oxygen concentration alarm setting, 1 setting the alarm values (individual settings) – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 159

Advertising
background image

IM 11M12D01-01E

9-9

9. Calibration

Table 9.1 When Auto Calibration conflicts with Other Commands

T9.1E.eps

X : Cannot perform calibration
Wait : Wait until other command finishes before starting calibration

Purging

the

channel

Other

channel

CAL

Indication

Check in

other channel

Blowback in

other

channel

CAL chan.
in warmup

status

CAL chan.

in ERR

status

CAL chan.

now

disabled

Maintenance

Man. Cal.

SemiAuto

AutoCal

Wait

Wait

Wait

Wait

Wait

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x
x

x

x

x

Other
commands

<Shipping-time defaults>
At shipping-time (or after resetting parameters) default values are as shown in Table 9.2
below.

Table 9.2

T9.2E.eps

Item

Initial value

Mode Manual

Points Span-Zero

Zero gas conc.

1.00%

Span gas conc.

21.00%

Hold time

3 min. 00 sec.

Cal. time

3 min. 00 sec.

Interval

1 days 00 hrs

Starting date

yy01mm01dd01

Starting time

hh00mm00