Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 21

Advertising
background image

IM 11M12D01-01E

2-2

(Excluding the case where the reference air flow is natural convection)

(Use oxygen of known concentration (in the measuring range) as the zero and

span calibration gas.)

Ϯ1% Maximum value of range setting ; from 0 to 5 vol% O

2

to 0 to 25

vol% O

2

range

(Sample gas pressure: within

Ϯ4.9 kPa)

Ϯ3% Maximum value of range setting ; from 0 to 25 vol%O

2

to 0 to 50 vol%

O

2

range

(Sample gas pressure: within

Ϯ0.49 kPa)

Ϯ5% Maximum value of range setting ; from 0 to 50 vol% O

2

to 0 to 100 vol%

O

2

range (Sample gas pressure: within

Ϯ0.49 kPa)

Drift: (Excluding the first two weeks in use)

(Excluding the case where the reference air flow is natural convection.)

Both zero and span

Ϯ2% Maximum value of range setting/month

Response Time : Response of 90% within 5 seconds. (Measured from when gas is

introduced from calibration gas inlet and analog output start changing.)