2 piping for the calibration gases – Yokogawa Multi Channel Oxygen Analyzer System ZR22/AV550G User Manual

Page 69

Advertising
background image

IM 11M12D01-01E

4-3

4. Piping

4.1.1

Parts Required for Piping in a System Using Flow Setting Units for Manual
Calibration

Referring to Table 4.1, check that the parts required for your system are ready.

Table 4.1

Detector

Piping Location

Part

Description

General-purpose Calibration gas inlet Stop valve

Yokogawa recommended:

detector

L9852CB or G7016XH

* Nipple

Rc1/4 or 1/4 NPT

General

Tube fitting Rc1/4 (1/4NPT) for a

␾6 ϫ␾4mm

General

soft tube

Reference gas inlet

(sealed up) (when piping is required, refer to § 4.1.3)

High-temperature Calibration gas inlet

Stop valve

Yokogawa recommended:

detector

L9852CB or G7016XH

* Nipple

Rc1/4 or 1/4 NPT

General

Tube fitting Rc1/4 (1/4 NPT) for a

␾6 ϫ␾4mm soft tube

Reference gas inlet

(sealed up)

(when piping is required, refer to § 4.1.3)

Sample gas outlet

* Auxiliary ejector

Yokogawa recommended,

E7046EC or E7046EN

* Equal tee, connection R1/4 or 1/4 NPT

General

fitting

* Needle valve

Rc1/4 or 1/4 NPT

General

* Reducing nipple R1/2 to R1/4 or R1/2 to 1/4 NPT

General

Note:

Parts marked with an asterisk (*) are used when required.
General parts are commercially available.

T4.1E.EP

S

4.1.2

Piping for the Calibration Gases

This piping is to be installed between the zero gas cylinder and the ZA8F flow setting
unit, and between the ZA8F flow setting unit and the ZR22G detector.

The cylinder should be placed in a calibration gas unit case or the like to avoid any
direct sunlight or radiant heat so that the gas cylinder temperature does not exceed 40

ЊC.

Mount a regulator valve (specified by YOKOGAWA) on the cylinder.

Mount a check valve or stop valve (specified by YOKOGAWA) on the nipple (found on
the open market) at the calibration gas inlet of the detector as illustrated in Figure 4.8.
(The check valve or the stop valve may have been mounted on the detector when
shipped.) Connect the flow setting unit and the detector to a stainless steel pipe 6 mm
(O.D.) ~ 4 mm or larger (I.D.) (or nominal size 1/4 inch).

F4.2E.EPS

Nipple

Stop valve

Tube connection
joint

Figure 4.2 Connection to the calibration gas inlet