4. Piping

4.2.7

Piping to Introduce Purge Gas When a Process Gas Alarm Occurs

When a process gas alarm (an input contact signal of unburnt gas detection) occurs, the
averaging converter will cut off the power supply to the heater of the detector, and at the
same time it will send a contact output to activate a solenoid valve for introducing a
purge gas to the detector.

The piping for purging in a system to perform automatic calibration is the same as the
one in the system using flow setting units for manual calibration. As instructed in
Subsection 4.1.7, this piping also requires that a check valve be installed on the
calibration gas inlet of the detector. A typical piping diagram for purging is shown in
Figure 4.13.

It is recommended that each instrument be installed to allow for minimum piping
between the solenoid valve for introducing the calibration gases and the check valve,
and between the solenoid valve for introducing the purge gas and the check valve.

To reference gas flow setting
flowmeter and needlre valve.

To reference gas flow setting flowmeter and needle valve.

Calibration contact outputs

Check valve

Check valve

To calibration gas flow setting needle valve and flowmeter.

Calibration
gas line

Reference
gas line

To calibration gas flow setting
needle valve and flowmeter.

Averaging converter (AV550G)

Heater

Signal

Contact output

Purge gas cylinder

Flowmeter

Pressure regulator

Needle valve

Reference
gas line

Calibration
gas line

Solenoid valve

Detector

Power supply

Contact input

Contact output

Analog output

(Averaged and individual outputs)

Stop valve

ء

: If optional 24 V outputs are specified for solenoid valve (MS Code "/24"), no external power supply for solenoid valves is required.

The solenpid valves are powered from the AV550G Averaging Converter. Never connect external power sources in the wiring for
solenoid valvas.

ء

ء

Figure 4.13 Typical Piping for Purging